College Physics ›› 2023, Vol. 42 ›› Issue (1): 19-.doi: 10.16854/j.cnki.1000-0712.220131
Previous Articles Next Articles
NI Lei,XIA Bo-long,LOU Jian-ling,XU Jin-yan
Received:
Revised:
Online:
Published:
Abstract: The energy spectrum of a 60Co source is measured by a 1-meter long plastic scintillator with a highly integrated Pixie-net digitizer. The position of the long plastic scintillator is calibrated by using the half-height position of Compton platform of 60Co sources. It is found that the difference between the calibration results and the actual position of radiation source is about 1~2 cm. The experimental results also show that the Pixi-net digitizer has strong ability to process the amplitude of a signal, and can replace QDC in the experiment of position sensitive plastic scintillator spectrometer.
Key words: Pixie-net digitizer, plastic scintillator, compton scattering
NI Lei, XIA Bo-long, LOU Jian-ling, XU Jin-yan. Position sensitive plastic scintillator spectrometer based on Pixie-net digitizer[J].College Physics, 2023, 42(1): 19-.
0 / / Recommend
Add to citation manager EndNote|Reference Manager|ProCite|BibTeX|RefWorks
URL: https://dxwl.bnu.edu.cn/EN/10.16854/j.cnki.1000-0712.220131
https://dxwl.bnu.edu.cn/EN/Y2023/V42/I1/19
Cited