College Physics ›› 2025, Vol. 44 ›› Issue (3): 53-.doi: 10.16854/j.cnki.1000-0712.240198

Previous Articles     Next Articles

Refractiveindex measurement experimental system based on phase-sensitive  spectral interferometry

ZHOU Hong-xian, WANG Yi   

  1. Experimental Education Center, Northeastern University at Qinhuangdao, Qinhuangdao, Hebei 066004, China
  • Received:2024-04-25 Revised:2024-05-25 Online:2025-05-09 Published:2025-06-03

Abstract: A refractive index measurement method and an experimental system based on spectral domain interference have been established, which mainly includes a fiber-optic Michelson interferometer, a spectrometer, linear fitting and phase unwrapping methods. Interference signals are collected by the spectrometer, and high-precision optical path lengths are obtained using linear fitting and phase unwrapping methods, thereby calculating the refractive index. Using this system, the refractive indices of silicon wafers, N-BK7 glass, and distilled water are measured to be 3.511±0.012, 1.508±0.019, and 1.311±0.022, respectively. The relative errors are 0.23%, 0.33%, and 0.37%, respectively. This method enables refractive index measurements of transparent and translucent solid and liquid samples without the need for special sample treatment, showing a new development and application of interferometry technology.


Key words: interferometry, phase-sensitive spectral domain interference, refractive index, phase unwrapping